Chemstrip 101 data error: Q: I get an error when I click "get data from analyzer" with my Chemstrip 101 analyzer. It says "error in transmission no STX found". A: Make sure your Roche Chemstrip 101 is set to bi-directional printing. Press the following buttons on the analyzer: "MENU, SETUP1, MODE, INTERF, BIDIR"
|